[Standard Deviation from 10 readings]
The tables below show readings of hazardous elements in resin and stainless alloy measured by the SEA1200VX. SEA1200VX employs the Vortex high count rate detector has better repeatability and reproducibility than the conventional models. |
| |
(1) Cadmium (Cd) & lead (Pb) in Resin
|
|
<Measurement Conditions>
Meas. Time : |
100 sec / filter |
|
Voltage : |
50kV, 15kV(Cr) |
|
Current : |
Auto |
Method : |
Calibration Method |
|
| |
|
Conventional Model |
SEA1200VX |
| PE Sample (Each Conc is 50ppm) |
Cd |
3.4 |
1.9 |
Pb |
0.7 |
0.5 |
Cr |
2.6 |
0.7 |
PVC Sample (Each Conc is 50ppm) |
Cd |
4.0 |
2.8 |
Pb |
3.1 |
1.8 |
Cr |
7.5 |
2.5 |
| Resin with Flame Retardant (Each has 200ppm) |
Cd |
9.0 |
3.5 |
Pb |
21.0 |
4.8 |
|
| |
(2) Readings from SUS |
|
| |
<Measurement Conditions>
Meas. Time : |
100 sec (1 condition) |
Voltage : |
50kV |
Current : |
Auto |
Method : |
Non-Standard Bulk FP |
|
| |
Conventional Model |
SEA1200VX |
| |
Ave. |
STDEV |
Ave. |
STDEV |
| Fe |
72.28 |
0.065 |
72.51 |
0.023 |
| Cr |
18.63 |
0.059 |
18.80 |
0.008 |
Ni |
7.85 |
0.043 |
7.36 |
0.017 |
| Mn |
1.18 |
0.055 |
1.27 |
0.008 |
| Mo |
0.05 |
0.007 |
0.06 |
0.001 |
|
|