Measurement Example of Environmentally Restricted Substances

Measurement of Brass Sample

High sensitive measurement conducted with SEA1200VX enables significantly better “limit of detection” and superior repeatability.

[ Measurement of lead (Pb) in Brass]

Excellent linearity of a calibration curve for brass created at optimal measurement conditions achieves a detection lower limit of Pb 16 ppm at a measuring time of 300 seconds.
Measurement of Pb in Brass
Measurement of Pb in Brass
(Brass sample made by Sumitomo Metal Technology)

[ Measurement of cadmium (Cd) in Brass]

Excellent linearity of the calibration curve was also obtained at optimal measurement conditions resulting in an excellent detection lower limit of 5 ppm Cd at a measuring time of 300 seconds.
Measurement of Cd in Brass
Measurement of Cd in Brass
(Brass sample made by Sumitomo Metal Technology )

[ Measurement of chromium (Cr) in Brass]

An superior detection lower limit of 18ppm at 300 seconds was achieved at optimal measurement conditions.
Measurement of Cr in Brass
Measurement of Cr in Brass
(Brass sample made by Sumitomo Metal Technology )

[ Repeatability at 100 seconds measurement]

The table below shows high repeatability using a brass standard sample: CV 6.3% for Cd, CV 3.6% for Pb, and CV 9.0% for Cr.

  < Measurement Conditions >
Voltage : 50kV(Cd, Pb), 15kV(Cr)
Current : 1000μA(Cd), 273μA(Pb), 353μA(Cr)
Collimator : φ8mm
Filter : for Cd, for Pb, & for Cr
Time : 100 sec

< Standard Sample >
Brass GBR6 made by Sumitomo Metal Technology
 
Cd
Pb
Cr
Ave. (ppm)
129
492
149
Readings
20
STDEV (ppm)
8.2
17.7
13.5
CV%
6.3
3.6
9.0
Results (ppm)
136
505
157