[ Repeatability at 100 seconds measurement]
The table below shows high repeatability using a brass standard sample: CV 6.3% for Cd, CV 3.6% for Pb, and CV 9.0% for Cr.
| |
< Measurement Conditions >
| Voltage : |
50kV(Cd, Pb), 15kV(Cr) |
| Current : |
1000μA(Cd), 273μA(Pb), 353μA(Cr) |
| Collimator : |
φ8mm |
| Filter : |
for Cd, for Pb, & for Cr |
| Time : |
100 sec |
< Standard Sample >
Brass GBR6 made by Sumitomo Metal Technology |
| |
Cd |
Pb |
Cr |
| Ave. (ppm) |
129 |
492 |
149 |
| Readings |
20 |
| STDEV (ppm) |
8.2 |
17.7 |
13.5 |
| CV% |
6.3 |
3.6 |
9.0 |
| Results (ppm) |
136 |
505 |
157 |
|
|