[ Measurement of Polyethylene Resin Standard ]
SII NanoTechnology developed a filter for measuring chlorine. A lower detection limit of 20ppm is achieved even in air, which is an advantage to vacuum since vacuum draw-down time is not required. Liquid and powder samples can also be measured.
•Capable of Cl measurement under "Air Condition"
•Capable of measuring Cl in liquids and powders
•Excellent DLL (in PE) : 20ppm |
< Measurement Condition (SEA1200VX) >
| Meas. Time(sec): |
100 |
Filter : |
for Cl
|
| Collimator : |
φ8.0mm |
Mylar : |
OFF |
| Voltage(kV) : |
15 |
Atmosphere : |
Air |
| Current(µA): |
545 |
Peaking Time : |
1.0usec |
|
< Measurement of Cl in BCR680 >
Certified Value 810 ± 16
| |
Cl Concentration
(ppm) |
| # of measurement |
10 times |
| Ave. |
873 |
| Range |
34 |
| STDEV |
11.4 |
| CV(%) |
1.3 |
|
 |
|